|  | MI 34 H with 1000 Hz High Frequency Probe Tone | |  | Fast, automatic tests - Tympanogram in 3 seconds | |  | Ipsi- and contralateral reflex tests with 4 frequencies | |  | Reflex Decay Test | |  | Eustachian tube function test for intact and perforated eardrums | |  | Handheld, adjustable probe for diagnostic and screening purposes | |  | Graphical realtime display | |  | Fast, virtually silent printout of test results | |
 | | Probe frequency, intensity | 226 Hz ± 1%, 85 dB SPL into 2 cm3 |  |
 | | High frequency (MI 34 H) | 1 kHz ± 1% |  |
 | | Pressure range | + 200 to - 400 daPa |  |
 | | Volume range | 0,1 to 6,0 ml |  |
 | | Accuracy | ± 5% or ± 10 daPa |  |
 | | Test time | < 3 seconds |  |
 | | Test frequencies | 500, 1000, 2000, 4000 Hz ± 2% |  |
 | | Test method | ipsilateral, contralateral |  |
 | | Intensities ipsilateral | 70 ... 105 dB HL |  |
 | | Intensities contralateral | 70 ... 120 dB HL with TDH 39 contra phone |  |
 | | Ipsilateral reflex test | with AGC |  |
 | | Intensity setting | automatic or manual |  |
 | | Test time | 13 sec., 10 sec. auto tone present |  |
 | | Pressure range | + 300 to - 400 daPa |  |
 | | Standard | IEC 601-1,IEC 645-5 class 1, medical directive 93/42/EEC, CE 0124 |  |
 | | Test program | Reflex test selectable |  |
 | | Memory | Storage of test results for both ears |  |
 | | Menu | For adapting to individual needs (Pump speed, accuracy, reflex mode etc.) |  |
 | | Probe | Lightweight adjustable hand-held probe with built-in control light and switch |  |
 | | Printer | fast, virtually silent thermal printer |  |
 | | Print time | 4 s / 12 s (Test results of both ears) |  |
 | | Display | Graphic LCD-Display, contrast adjustable |  |
 | | PC-interface | USB |  |
 | | Power supply | Mains 100 ... 240 V~, 50/60 Hz, 25 VA |  |
 | | Dimension and weight | W x D x H: 39 x 29 x 11 cm / 2,6 kg |  |
| | |
| |