|  | Ready to use via USB connection to a PC | |  | Small and light, flexible and highly mobile | |  | Keyboard or mouse controlled volume and frequency range operation | |  | Air and Bone Conduction, Speech and SISI Test available | |  | No disposable material | |  | Results immediately displayed and stored | |  | Including Patient database | |
 | | Test Signal | Pure, Pulse or Warble tone |  |
 | | Test Frequency | 0,125; 0,25; 0,5; 0,75; 1; 1,5; 2; 3; 4; 6; 8 kHz |  |
 | | Intensity | 5 dB noise free |  |
 | | Audiometer Class | Class 4 according to EN 60645-1 |  |
 | | Masking | with narrow band noise masking, automatic or manually adjustable |  |
 | | Safety Standards | EN 60601-1, EN 60645-1 Instrument complies with 93/42/EWG Directive for Medical Product Class II a. |  |
 | | Standard accessories | MA 33 box; air conduction headphone, patient response key, Tone audiometric software, Patient database, USB cable, hand-carry bag |  |
 | | Test intensity | from -10 to 110 dBHL Depending on headphone and frequency |  |
 | Air conduction Air conduction headphone | HDA 280 , TDH 39, DD 45 or Holmco 8103 |  |
| | | Bone Conduction (MA 33 BC / MA 33 SISI / MA 33 Speech) |
 | | Test intensity | -10 bis 70 dBHV (40 dB at 6 kHz, 60 dB at 0,5 and 4 kHz, 65 dB at 3 kHz) |  |
 | | Bone conduction headphone | B71 with headband |  |
 | | Modulation Test signal | 4,8/0,2 s; 5 dB, 3 dB, 2 dB (preparation) 4,8/0,2 s; 1 dB (Test) |  |
 | | Power supply | via USB |  |
 | | Size and weight | W x D x H: 15,4 x 10,3 x 2,7cm / 300 g |  |
 | | Computer requirements | 1.1 or 2.0 USB port, Windows XP SP2, Intel P4 > 1 GHz or compatible, 1 GB RAM |  |
| | |
| |